COOK INC DOUBLE LUMEN PERIPHERALLY INSERTED CENTRAL VENOUS CATHETER SET; FOZ CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM LESS THAN 30 DAYS Back to Search Results
Catalog Number PICDS-601-MPIS-NT
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/25/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported a doctor noticed the double lumen peripherally inserted central venous catheter set was leaking thirteen days after picc placement at the large lumen after he had clamped the line proximal to the hole.As reported, the doctor assumes the clamp caused this leak, however, the device remains in the patient as he is still using the small lumen of by the large lumen of the catheter.As reported, the patient did not experience any adverse effects or require any additional procedures due to this occurrence.
 
Additional Manufacturer Narrative
Corrected data:g5-pma/510(k) # ;h10: the leak was discovered approximately 9 days after implant, not 13 days as previously reported.Additional information: the customer stated the patient had no pre-existing condition and/or pre-diagnosis relevant to this event.Investigation - evaluation: a review of the complaint history, device history record, documentation, drawing, manufacturing instructions, quality control, and specifications of the device was conducted during the investigation.The complaint device was not returned; therefore, no physical examinations could be performed.However, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record shows no nonconforming events which could contribute to this failure mode.It should be noted there were no other reported complaints for this lot number.Based on the information provided, no product returned and the results of our investigation, a definitive root cause could not be determined.We will continue our monitoring of similar complaints and have notified the appropriate personnel of this event.Per the quality engineering risk assessment no further action is required.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DOUBLE LUMEN PERIPHERALLY INSERTED CENTRAL VENOUS CATHETER SET
Common Device Name
FOZ CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM LESS THAN 30 DAYS
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7444755
Report Number1820334-2018-01047
Device Sequence Number11256953
Product Code FOZ
UDI-Device Identifier00827002086654
UDI-Public(01)00827002086654(17)190217(10)6743154
Combination Product (Y/N)N
Initial Reporter CountryGB
PMA/510(K) Number
K950118
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberPICDS-601-MPIS-NT
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/27/2018
Supplement Date Received by Manufacturer05/21/2018
Initial Report FDA Received Date04/19/2018
Supplement Report FDA Received Date05/24/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-