It was reported a doctor noticed the double lumen peripherally inserted central venous catheter set was leaking thirteen days after picc placement at the large lumen after he had clamped the line proximal to the hole.As reported, the doctor assumes the clamp caused this leak, however, the device remains in the patient as he is still using the small lumen of by the large lumen of the catheter.As reported, the patient did not experience any adverse effects or require any additional procedures due to this occurrence.
|
|
Corrected data:g5-pma/510(k) # ;h10: the leak was discovered approximately 9 days after implant, not 13 days as previously reported.Additional information: the customer stated the patient had no pre-existing condition and/or pre-diagnosis relevant to this event.Investigation - evaluation: a review of the complaint history, device history record, documentation, drawing, manufacturing instructions, quality control, and specifications of the device was conducted during the investigation.The complaint device was not returned; therefore, no physical examinations could be performed.However, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record shows no nonconforming events which could contribute to this failure mode.It should be noted there were no other reported complaints for this lot number.Based on the information provided, no product returned and the results of our investigation, a definitive root cause could not be determined.We will continue our monitoring of similar complaints and have notified the appropriate personnel of this event.Per the quality engineering risk assessment no further action is required.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|