Model Number S7 |
Device Problems
Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138); Device Operates Differently Than Expected (2913)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 03/29/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
No patient information provided as no patient was involved in this concern.A medtronic representative went to site to test the equipment.Representative reported that the computer would sometime not bootup.Replacing the computer resolve the issue.A system checkout was performed and the hardware, software, and instruments passed the system checkout.The system was found to be fully functional.The suspect computer has not been received by the manufacturer for evaluation.
|
|
Event Description
|
Medtronic received information that while outside of procedure, the navigation system would unexpectedly become unresponsive, shutdown and reboot with data loss.There was no patient present at the time of the issue.
|
|
Manufacturer Narrative
|
The computer for the navigation system was returned to the manufacturer for analysis.The computer was found to be fully functional with no problem found.The reported event could not be duplicated by medtronic personnel.
|
|