COVIDIEN ARGYLE; CATHETER, UMBILICAL ARTERY Back to Search Results
Model Number 8888160333
Device Problem Leak/Splash (1354)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Submit date: 4/19/2018.The incident sample has been requested but to date has not been received for evaluation.If the sample is received, or if additional information pertinent to the incident is obtained a follow-up report will be submitted.As part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.
 
Event Description
The customer reports cracks have been observed.
 
Manufacturer Narrative
An investigation was performed for the reported customer complaint: ¿the customer reports cracks have been observed.¿ the device history record (dhr) for the reported lot number 1523900126 indicates product and specification requirements were met with no non-conforming product identified relating to this customer report.A lot cannot be released unless it passes all quality and conformance requirements.In-process procedures are also in place to prevent nonconforming product in the manufacturing process.This ensures components and finished products meet all quality inspection standards.These controls include, but are not limited to: material verification/certification processes, dimensional specifications, statistical samplings, periodic audits, process inspections, machine maintenance/operation and personnel training and certification.During the manufacturing process, catheters are submitted to 100% pressure testing which would identify this reported issue in the catheter assembly.Based on available information, it can be concluded the product was manufactured according to specifications.No product/sample was provided for evaluation.No additional information, pictures or videos were received.Therefore, a comprehensive investigation was unable to be conducted.The reported customer complaint is unable to be confirmed.A root cause could not be determined.This complaint will be utilized for tracking and trending purposes.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARGYLE
Type of Device
CATHETER, UMBILICAL ARTERY
Manufacturer (Section D)
COVIDIEN
edificio b20, calle #2
alajuela 20101
Manufacturer (Section G)
COVIDIEN
edificio b20, calle #2
alajuela 20101
Manufacturer Contact
edward almeida
15 hampshire street
mansfield, MA 02048
5084524151
MDR Report Key7445052
MDR Text Key106026130
Report Number3009211636-2018-00120
Device Sequence Number1
Product Code FOS
UDI-Device Identifier20884527005113
UDI-Public20884527005113
Combination Product (y/n)N
Reporter Country CodeBE
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/27/2020
Device Model Number8888160333
Device Catalogue Number8888160333
Device Lot Number1523900126
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/26/2018
Initial Date FDA Received04/19/2018
Supplement Dates Manufacturer Received03/26/2018
Supplement Dates FDA Received09/19/2018
Date Device Manufactured08/27/2015
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
-
-