| Catalog Number |
CAT5 |
| Medical Device Problem Codes |
Bent (1059); Failure to Advance (2524)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/23/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Results: the cat5 was fractured approximately 107.0 cm from the hub.Conclusions: evaluation of the returned cat5 revealed a fracture on its distal shaft.If the device is advanced against resistance at extreme angles while introducing the device into a parent catheter, a kink may occur.If the damaged device is further mishandled it may result in the eventual fracture of the catheter.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
| |
|
Event or Problem Description
|
|
The patient was undergoing a thrombectomy procedure using an indigo system cat5 aspiration catheter (cat5).During the procedure, while attempting to insert a cat5 into a non-penumbra sheath, the cat5 would not advance and subsequently, became bent; therefore, it was removed.The procedure was completed using the same sheath and an indigo system cat3 aspiration catheter (cat3).There was no report of an adverse effect to the patient.
|
| |
|