• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS PROV 4F DL IR 135CM NI; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS PROV 4F DL IR 135CM NI; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Catalog Number S3274335
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) review is not possible, as no manufacturing lot number has been provided by the complainant.(b)(4).
 
Event or Problem Description
Per medwatch; reportedly, called to assess picc line because of a report that it has a tear in the lumen.When i looked at picc line the purple hub and cap was only barely attached to the picc lumen.When i turned the purple cap to assess it, the purple line came off in my hand.The picc line was then removed and nurse updated.
 
Additional Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, frequency analysis, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the complaint of a break is confirmed and corrective action has been taken regarding failures of this type.One 4 fr dual lumen power picc provena was returned for evaluation.An initial visual observation showed the purple luer hub was missing and use residue was observed throughout the sample.The pinch clamp on the red lumen was observed to be engaged and the pinch clamp on the purple lumen was observed to be missing.The ink on both extension legs was observed to be worn and faded, and a needleless injection cap was observed to be attached to the red luer hub.A microscopic observation revealed the fracture surface of the extension tubing of the purple lumen was rough and exhibited cracks/fissures and beach marks.A lot history review (lhr) review is not possible, as no manufacturing lot number has been provided by the complainant.
 
Event or Problem Description
Per medwatch; reportedly, called to assess picc line because of a report that it has a tear in the lumen.When i looked at picc line the purple hub and cap was only barely attached to the picc lumen.When i turned the purple cap to assess it, the purple line came off in my hand.The pic c line was then removed and nurse updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROV 4F DL IR 135CM NI
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key7445749
Report Number3006260740-2018-00764
Device Sequence Number107238
Product Code LJS
UDI-Device Identifier00801741130045
UDI-Public(01)00801741130045
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K162443
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type User Facility
Initial Reporter Occupation Other Health Care Professional
Remedial Action Recall,Other
Type of Report Initial,Followup
Report Date (Section B) 02/14/2019
Report Date (Section F) 02/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberS3274335
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/18/2019
Was the Report Sent to FDA? Yes
Event Location Hospital
Date Report to Manufacturer03/23/2018
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer02/06/2019
Initial Report FDA Received Date04/19/2018
Supplement Report FDA Received Date02/14/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age74 YR
Date Report Sent to FDA02/01/2018
-
-