| Model Number |
BA22-80/I16-40 |
| Medical Device Problem Codes |
Failure To Adhere Or Bond (1031); Hole In Material (1293); Leak/Splash (1354)
|
| Health Effect - Clinical Codes |
Aneurysm (1708); Failure of Implant (1924)
|
| Date of Event |
12/11/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The devices involved in this event will not be returned for evaluation and remain implanted in the patient.If additional information is obtained at a later time that is pertinent to this event, a follow-up report will then be submitted.
|
| |
|
Event or Problem Description
|
|
The patient was implanted with an afx bifurcated stent graft and vela suprarenal on (b)(6) 2015.On a later date, the physician suspected a type 3a endoleak, but this leak was not confirmed.In addition, there was possible billowing and a type 2 endoleak.No adverse events were reported for this patient whom was asymptomatic.No devices were explanted as the physician relined the afx system with an afx2 bifurcated stent graft and new vela suprarenal.
|
| |
|
Additional Manufacturer Narrative
|
|
At the completion of the clinical evaluation and based on the information received, there was substantial evidence to support the following reported events: endoleaks 1a and 2, and endovascular relining.There was substantial evidence to refute the following reported events: endoleak 3a, rather there was an endoleak 3b noted one month post-implant (left lateral distal near bifurcation).Additional findings include sac growth (8mm/27m) at 27 months post-implant.The most likely cause of the proximal loss of seal was the hostile aortic neck anatomy and the presence of a patent, low accessory renal artery.The most likely cause of the compromised stent graft integrity (breached) of the main body stent was the use of strata material.Procedure-related harms included a type 2 endoleak.The final patient disposition could not be determined due to the lack of medical information surrounding the implant and repair events.The manufacturing lot review confirmed all devices met specifications prior to release.Endologix continues to investigate this event and similar events to ensure the highest quality and patient safety.A root cause investigation was carried out for all afx complaints having an identified failure mode of a type 3b endoleak.Endologix implemented the following corrective actions with the intent of reducing type 3b endoleak events; 1.Upgraded graft material (i.E.Duraply) and 2.Updates to the ifu and additional physician training.The change to duraply graft material and the ifu changes were put in place july 2014.The type 3b endoleak rate for afx manufactured and implanted before these corrective actions were put in place is trending at 2.5%.Since the corrective actions were implemented, the type 3b endoleak events reported for afx devices has been reduced to <0.2%.Patient codes: add (b)(4).Device codes: add (b)(4).Remove (b)(4).Method codes: add (b)(4).- (b)(4).Conclusion codes: add 22.
|
| |
|
Event or Problem Description
|
|
Clarification: endologix became aware of a re-intervention on (b)(6)2017 as reported by physician linwood bosher at bon secours memorial regional.As the date of event is unknown, the best estimate (b)(6)2017 was used.The patient was originally implanted with an afx bifurcated stent graft and vela suprarenal on (b)(6)2015 by the reporting physician.On an unknown date, the physician suspected a 3a endoleak, which was not confirmed, and possible billowing with a type 2 endoleak.The patient was asymptomatic.During re-intervention, the patient was relined by implanting an afx2 bifurcated stent graft and one additional vela suprarenal on (b)(6)2017 by physician linwood bosher.
|
| |
|