ENDOLOGIX INC. AFX; BIFURCATED STENT GRAFT Back to Search Results
Model Number BA25-90/I20-30
Medical Device Problem Codes Failure To Adhere Or Bond (1031); Leak/Splash (1354)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 02/01/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The devices involved in this event will not be returned for evaluation and remain implanted in the patient.If additional information is obtained at a later time that is pertinent to this event, a follow-up report will then be submitted.
 
Event or Problem Description
The patient was implanted with an afx bifurcated stent graft and a veal suprarenal on (b)(6) 2015.Contrast was observed at a routine follow-up ct scan, at which time a type 3a endoleak was observed.The patient was asymptomatic.A vela infrarenal was added on (b)(6) 2016, immediately after which the endoleak was eliminated and the patient recovered well.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFX
Common Device Name
BIFURCATED STENT GRAFT
Manufacturer (Section D)
ENDOLOGIX INC.
2 musick
irvine CA 92618
MDR Report Key7445914
Report Number2031527-2018-00191
Device Sequence Number19113819
Product Code MIH
UDI-Device Identifier00818009012227
UDI-Public(01)00818009012227(17)180706
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
P040002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Reporter Type Manufacturer
Report Source company representative,other
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date07/06/2018
Device Model NumberBA25-90/I20-30
Device Lot Number1293160-014
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/01/2016
Initial Report FDA Received Date04/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AFX-VELA SUPRARENAL - LOT 1251980-008
Outcome Attributed to Adverse Event Required Intervention;
Patient Age68 YR
-
-