| Model Number |
BA25-90/I20-30 |
| Medical Device Problem Codes |
Failure To Adhere Or Bond (1031); Leak/Splash (1354)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
02/01/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The devices involved in this event will not be returned for evaluation and remain implanted in the patient.If additional information is obtained at a later time that is pertinent to this event, a follow-up report will then be submitted.
|
| |
|
Event or Problem Description
|
|
The patient was implanted with an afx bifurcated stent graft and a veal suprarenal on (b)(6) 2015.Contrast was observed at a routine follow-up ct scan, at which time a type 3a endoleak was observed.The patient was asymptomatic.A vela infrarenal was added on (b)(6) 2016, immediately after which the endoleak was eliminated and the patient recovered well.
|
| |
|