| Model Number |
ZA9003 |
| Medical Device Problem Code |
Bent (1059)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
01/22/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The lens was inserted and removed.(b)(4).All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that after an intraocular lens (model za9003 +13.5 diopter) was inserted into the patient's eye, the surgeon noticed that the trailing haptic was bent close to the optic.Reportedly, the lens was removed and replaced with another lens (same model and diopter) during the same surgical procedure.The incision was enlarged to remove the lens but no sutures were used.No further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Device available for evaluation? yes.Returned to manufacturer on: 04/30/2018.Device returned to manufacturer? yes.Device evaluation: the intraocular lens (iol) was returned at the manufacturing site for evaluation.Visual inspection showed that both haptics were distorted.Traces of what appeared to be ovd (opthalmic viscosurgical device) were observed on the lens surface.The customer's reported complaint was verified.Since the lens was handled at the surgical site it could not be possible to determined that the damages were related to the manufacturing process.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search revealed that no additional complaints for this order number have been received.Labeling review: the directions for use (dfu) were reviewed.The directions for use (dfu) adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
|
| |
|
Search Alerts/Recalls
|