|
|
| Catalog Number |
RBY2C0520 |
| Medical Device Problem Code |
Detachment Of Device Component (1104)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/21/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The pet lock was intact on the proximal end of the pusher assembly.The pusher assembly was kinked approximately 74.0 cm from the proximal end.The stretch resistant wire (sr wire) was fractured, the proximal constraint sphere was missing from the distal detachment tip (ddt) and the embolization coil was detached from its pusher assembly.Evaluation of the first evaluated ruby coil revealed that the sr wire was fractured.This type of damage typically occurs due to improper handling during use.If the device is forcefully retracted against resistance, damage such as this may occur.Further evaluation revealed the sphere was missing from the distal detachment tip (ddt).This may have been due to the pusher assembly elongating which would allow the pull wire to retract from the ddt.However, since the sr wire is fracture this likely occurred after the coil detachment.The root cause of resistance could not be determined.Evaluation of the second evaluated ruby coil revealed that the pusher assembly was fractured.If the device is forcefully advanced against resistance, damage such as a kink may occur.If the kink is further manipulated or straightened, damage such as a fracture may occur.The root cause of resistance could not be determined.The non-penumbra microcatheter referred to in the complaint was not returned for evaluation.Penumbra coils are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report number: 3005168196-2018-00818.
|
| |
|
Event or Problem Description
|
|
The patient was undergoing a coil embolization procedure in the right middle lobe segment of the pulmonary artery using ruby coils.During the procedure, a ruby coil prematurely detached within the microcatheter, and therefore the microcatheter was removed with the ruby coil inside.The physician then inserted a new non-penumbra microcatheter, and attempted to place another ruby coil, however the ruby coil was unable to deploy.The ruby coil was therefore removed, and the procedure was completed using the same microcatheter and additional ruby coils.Although there was no report of an adverse effect on the patient, multiple attempts have been made to gather additional information for the second ruby coil.As of (b)(6) 2018, no additional information is available.
|
| |
|
Search Alerts/Recalls
|
|
|