Catalog Number 01.18.234
Health Effect - Clinical Code No Code Available (3191)
Date of Event 03/23/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Batch review performed on 17 april 2018; lot 172626: (b)(4) items manufactured and released on 11 october 2017.Expiration date: 2022-10-01; no anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold without any similar reported event.Preliminary investigation performed by r&d product manager on 19 april 2018: evaluating the images attached to this complaint it seems that the broach is functionally conform.There are only some little scratches visible on the broach surface, due to the usage, that we can consider acceptable.There is no evidence related to the fact that the surgeon wasn't able to insert the stem size 4 in the femur after canal preparation.Visual inspection performed by r&d product manager on 20 april 2018: the stem appeared without any particular sign, except for the scratches related to the difficult removal (signs under the collar, groove in the neck).From the received piece it is not possible to determine the root cause of the event.
 
Event or Problem Description
After having used the broach size #4, the surgeon tried to implant the stem size #4, but it floated about 7mm up.The surgeon replaced it with a stem size #3.It took about 40 minutes for retrieving.The broach was not defective.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEM COLLARED HA COATED STD STEM SIZE 4
Common Device Name
CEMENTLESS HIP STEM
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
SZ  6874
MDR Report Key7446376
Report Number3005180920-2018-00257
Device Sequence Number1950102
Product Code LZO
UDI-Device Identifier07630030804595
UDI-Public07630030804595
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K121011
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/01/2022
Device Catalogue Number01.18.234
Device Lot Number172626
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/16/2018
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/23/2018
Initial Report FDA Received Date04/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age40 YR
-
-