| Model Number |
24950JLQ |
| Medical Device Problem Codes |
Device Emits Odor (1425); Overheating of Device (1437)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/18/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the remote monitor plug caught on fire while plugged into a receptacle at the patient's home.The monitor is expected to be replaced.No patient complications have been reported as a result of this event.
|
| |
|
Search Alerts/Recalls
|