• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT ARMADA 35 PTA CATHETER; PERIPHERAL DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT ARMADA 35 PTA CATHETER; PERIPHERAL DILATATION CATHETER Back to Search Results
Catalog Number B1070-100
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/28/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.Guide wire: terumo.Sheath: fortress 6 fr crossover.Evaluation summary: visual analysis was performed on the returned device.The reported balloon rupture and separation were confirmed.The difficulty removing was not tested due to the condition of the device.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents from this lot.The investigation determined that the reported difficulties and subsequent patient effects were due to case circumstances.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event or Problem Description
It was reported that the procedure was to treat a de novo lesion located in the mid femoral artery that was mildly tortuous, heavily calcified and 80% stenosed.The armada 35 percutaneous transluminal angioplasty catheter was advanced with success however the balloon ruptured during the first inflation at 8 atmospheres due to excessive calcification in the treatment area.During removal of the delivery system with slight resistance, it was observed that 80 cm came out of the sheath in the patient and the proximal 10-12 cm of the system was still in crossover position in the treated area.The delivery system broke in two pieces in the patient.A snare device was used to retrieve the separated segment.This was done with success and no adverse patient effects and it was confirmed nothing was left in the patient.There was a clinically significant delay of 90 minutes and an extra 5000 units of heparin was given to patient.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 35 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7447182
Report Number2024168-2018-02925
Device Sequence Number1768511
Product Code LIT
UDI-Device Identifier08717648154812
UDI-Public08717648154812
Combination Product (Y/N)N
Initial Reporter CountryBE
PMA/510(K) Number
K111899
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 04/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2020
Device Catalogue NumberB1070-100
Device Lot Number70623G1
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/28/2018
Initial Report FDA Received Date04/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-