The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device has not been returned, at this time, to the manufacturer for evaluation.A lot history review (lhr) of reby2344 showed no other similar product complaint(s) from this lot number.Device has not yet been returned for evaluation.
|
It was reported the inserting rn met subcutaneous resistance when inserting midline but knew he was going through scar tissue, so persisted.It was stated the rn stopped and pulled the catheter out, it was noted that approximately 2cm of the tip of the catheter was missing.The patient went for a ct and x-ray and it showed multiple implants and stents in the arm.The inserting nurse stated that he may have sheered the catheter on an implant or stent in the patient's arm.
|