COVIDIEN MFG DC BOULDER FORCE TRIVERSE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number FT3000
Medical Device Problem Code Self-Activation or Keying (1557)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 04/05/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, prior to a laparoscopic partial nephrectomy, the device was activating although the activation button was not being pressed at that time.A new ft3000 was opened and it worked fine.Foot switch was not being stepped on at that time of issue.No patient involvement.
 
Additional Manufacturer Narrative
Evaluation summary: one device sample was received for evaluation.This device had been used in the treatment or diagnosis of a patient.A review of the lot number reported indicates that the product was within the assigned expiration date at the time of the reported incident.The returned product met specification as received.Visual inspection found no defects.The investigation found the device to function normally and within specifications.The device was recognized when plugged into a generator.The device was not self activating when plugged into the generator.The device was activated five times in all three modes and all five slider positions with no anomalies noted.The investigation found the device to function normally and within specifications.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FORCE TRIVERSE
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
MDR Report Key7447320
Report Number1717344-2018-00528
Device Sequence Number11276121
Product Code GEI
UDI-Device Identifier20884524000081
UDI-Public20884524000081
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K150735
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/25/2022
Device Model NumberFT3000
Device Catalogue NumberFT3000
Device Lot Number72050165X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/30/2018
Initial Date Received by Manufacturer 04/06/2018
Supplement Date Received by Manufacturer08/09/2018
Initial Report FDA Received Date04/20/2018
Supplement Report FDA Received Date08/27/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-