BARD PERIPHERAL VASCULAR, INC. ATLAS PTA DILATATION CATHETER; PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number AT120144
Medical Device Problem Codes Peeled/Delaminated (1454); Retraction Problem (1536); Unraveled Material (1664)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/10/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No device and no medical records were provided to the manufacturer; however a photo of the sample was provided and is currently being investigated.The lot number for the device was provided.The device history records are currently under review.The investigation of the reported event is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that after a angioplasty procedure in the left iliac vein, the pta balloon fibers were found to be allegedly unraveled and peeling off of the balloon.It was further reported that there was difficulty retracting the balloon through the sheath.There was no reported patient injury.
 
Additional Manufacturer Narrative
The device history records were reviewed with special attention to the raw materials, subassemblies, manufacturing process and quality control testing.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported to date for this lot number and failure mode.The device was not returned, however one electronic photo was provided for review.Based on the photo review, the investigation is confirmed for peeling of the outer pebax.However, the investigation is inconclusive for the reported unraveled balloon fibers and retraction issue, as the photo review could not confirm the reported issues, and as the sample was not returned for full visual and functional evaluation of the device.The definitive root cause for the identified peeled pebax or reported fiber and retraction issues could not be determined based upon available information.The review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.(expiry date-07/2020).
 
Event or Problem Description
It was reported that after a angioplasty procedure in the left iliac vein, the pta balloon fibers were found to be allegedly unraveled and peeling off of the balloon.It was further reported that there was difficulty retracting the balloon through the sheath.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATLAS PTA DILATATION CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key7447442
Report Number2020394-2018-00420
Device Sequence Number105574
Product Code LIT
UDI-Device Identifier00801741062636
UDI-Public(01)00801741062636
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K120971
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date (Section B) 06/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/28/2020
Device Model NumberAT120144
Device Catalogue NumberAT120144
Device Lot NumberGFBU1447
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/26/2018
Supplement Date Received by Manufacturer06/07/2018
Initial Report FDA Received Date04/20/2018
Supplement Report FDA Received Date06/26/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-