| Model Number |
621-100ES |
| Medical Device Problem Code |
Thermal Decomposition of Device (1071)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/24/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The device involved is pending evaluation and the investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
|
| |
|
Event or Problem Description
|
|
As reported by the user facility: while infusing, the pump was found to have smoke coming from the pump itself and it was hot to the touch.The infusion was stopped.It was indicated there was a small amount of smoke observed at the back of the pump with a burnt electrical odor.Vancomycin 1 gram reconstituted in 250ml of 0.9% sodium chloride was infusing at the time.The volume to be infused was 250ml at a rate of 167ml/hr.There was no injury to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The device involved was received for evaluation.The incoming battery measured 1.6v.The failed 12v battery was replaced.The log review shows on (b)(6) 2018 and 9:23am a dc power-on and at 9:28am an infusion start of 167ml/hr and 264ml.At 9:28am a low battery warning occurred.On (b)(6) 2018 and 6:01am the pump powered-on with se 75, 123, indicating a previous unexpected pump power down.If additional pertinent information becomes available a follow-up report will be filed.
|
| |
|
Event or Problem Description
|
|
As reported by the user facility: while infusing, the pump was found to have smoke coming from the pump itself and it was hot to the touch.The infusion was stopped.It was indicated there was a small amount of smoke observed at the back of the pump with a burnt electrical odor.Vancomycin 1 gram reconstituted in 250ml of 0.9% sodium chloride was infusing at the time.The volume to be infused was 250ml at a rate of 167ml/hr.There was no injury to the patient.
|
| |
|