• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS Back to Search Results
Model Number MICL12.6
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Codes Patient Problem/Medical Problem (2688); No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
The reporter indicated that he implanted a 12.6 mm micl12.6, -10.5 diopter, implantable collamer lens into the patient's left eye (os) on (b)(6) 2017.On (b)(6) 2018 the lens was explanted due to chronic iritis.It was reported that the patient was treated with durezol and pred forte for over three months.The reported stated that in his opinion the cause of event was due to patient sulcus anatomy.On (b)(6) 2018 the surgeon reported that the patient is continuing a steroid taper and that the patient continues to recover.If additional information is received a supplemental medwatch report will be submitted.
 
Additional Manufacturer Narrative
Device evaluation: product evaluation found the lens returned dry in a cartridge case with residue on lens.Visual inspection found a piece of the haptic missing and residue on lens.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
MDR Report Key7448696
Report Number2023826-2018-00587
Device Sequence Number18868916
Product Code MTA
UDI-Device Identifier00841542103039
UDI-Public00841542103039
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date08/31/2019
Device Model NumberMICL12.6
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/08/2018
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 03/22/2018
Supplement Date Received by Manufacturer05/08/2018
Initial Report FDA Received Date04/20/2018
Supplement Report FDA Received Date06/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-