• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL, ASD, INC. FLEXTEND¿ TTS¿ TRACHEOSTOMY TUBE; TUBE TRACHEOSTOMY AND TUBE CUFF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL, ASD, INC. FLEXTEND¿ TTS¿ TRACHEOSTOMY TUBE; TUBE TRACHEOSTOMY AND TUBE CUFF Back to Search Results
Catalog Number ZT17LN80NSC098S
Medical Device Problem Codes Fluid/Blood Leak (1250); Leak/Splash (1354)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/24/2018
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a flextend¿ tts¿ tracheostomy tube was not holding saline.The defect was discovered after use with a patient.The product issue did not cause patient injury.
 
Additional Manufacturer Narrative
Two smiths medical portex bivona flextend¿ tts¿ tracheostomy tubes were returned for analysis.Visual exam performed showing no flaws.Leak tests were performed indicating that a small leak was noted in one of the tracheostomy (trach) tubes.The amount of adhesive surrounding the airway line at the neck-flange junction was noted to be adequate.Based on the evidence, the root cause of the leak was unable to be determined if it was manufacturing related or excessive force that would have been applied to the airway line.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXTEND¿ TTS¿ TRACHEOSTOMY TUBE
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
SMITHS MEDICAL, ASD, INC.
6000 nathan lane north
minneapolis MN 55442
MDR Report Key7448709
Report Number3012307300-2018-01222
Device Sequence Number1769116
Product Code JOH
UDI-Device Identifier15019517136029
UDI-Public15019517136029
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K923878
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/25/2023
Device Catalogue NumberZT17LN80NSC098S
Device Lot NumberDS012411
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/19/2018
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/29/2018
Supplement Date Received by Manufacturer05/02/2018
Initial Report FDA Received Date04/20/2018
Supplement Report FDA Received Date05/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age22 YR
-
-