• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN MELSUNGEN AG INTROCAN SAFETY®; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN MELSUNGEN AG INTROCAN SAFETY®; INTRODUCER, CATHETER Back to Search Results
Model Number 4252527-02
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Exemption number e2016018 b.Braun inc.(bbmi) (importer) is submitting the report on behalf of b.Braun melsungen ag (manufacturer).This report has been identified as b.Braun medical inc.Internal report number (b)(4).The investigation into this reported event is ongoing.Additional attempts to receive the sample are being made.A follow-up report will be submitted when the results of the investigation are available.
 
Event or Problem Description
As reported on medwatch (b)(4): the patient was transferred from a outside user facility.The rn had trouble flushing the iv line.Attempt to flush line in the right upper arm caused a lot of pain for the patient.The line could not be flushed.There was a hard plastic-like bulge felt where the catheter should be.The line was pulled and found to be very kinked.The tip seemed to be torn off.The provider was notified.A concern for the retained tip was noted and disclosure to the patient was done.Multiple diagnostic exams were performed to locate the tip of iv line.There was no verification by an exam that there was a retained foreign object.
 
Additional Manufacturer Narrative
Exemption number e2016018.(b)(4).The actual device involved in the reported incident was not returned for evaluation.Without the actual sample a thorough sample analysis could not be performed and no specific conclusions can be drawn.Review of the device history record performed for the reported lot number did not reveal any abnormalities or nonconformances of this nature.No adverse quality trends of this nature were identified during the complaint review process for the reported catalog number or lot number.If additional pertinent information becomes available a follow-up report will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTROCAN SAFETY®
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
B. BRAUN MELSUNGEN AG
carl-braun-str. 1
melsungen, hessen 34212
GM  34212
MDR Report Key7448735
Report Number9610825-2018-00054
Device Sequence Number1125174
Product Code DYB
UDI-Device Identifier04046963318185
UDI-Public(01)04046963318185
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K021094
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,other,use
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 04/25/2018
Report Date (Section F) 04/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/01/2021
Device Model Number4252527-02
Device Catalogue Number4252527-02
Device Lot Number16F13G8241
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date03/26/2018
Device Age21 MO
Event Location Hospital
Date Report to Manufacturer04/25/2018
Initial Date Received by Manufacturer 03/26/2018
Supplement Date Received by Manufacturer03/26/2018
Initial Report FDA Received Date04/20/2018
Supplement Report FDA Received Date04/25/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age77 YR
Date Report Sent to FDA04/25/2018
-
-