• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. RUBY COIL; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. RUBY COIL; HCG, KRD Back to Search Results
Catalog Number RBY2C1240
Medical Device Problem Codes Detachment Of Device Component (1104); Unraveled Material (1664)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/26/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.The device was implanted in the patient.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure using ruby coils.During the procedure, the physician placed the initial coils in the target vessel using a lantern delivery microcatheter (lantern).While attempting to deploy a ruby coil in the target vessel, the lantern backed up in the non-penumbra sheath and consequently, the ruby coil became tangled between the lantern and the sheath.Subsequently, the ruby coil unintentionally detached, with the left part of the ruby coil running down the left side of the gonadal vein.Therefore, the physician tried bringing a long sheath to the coil mass and push the rest of the ruby coil with a coil pusher and two balloons.However, the ruby coil became unraveled and remained in the patient's body.The procedure was ended at that point and there was no adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RUBY COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key7448810
Report Number3005168196-2018-00819
Device Sequence Number2053643
Product Code HCG
UDI-Device Identifier00814548013015
UDI-Public00814548013015
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K120330
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date06/25/2025
Device Catalogue NumberRBY2C1240
Device Lot NumberF77901
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/26/2018
Initial Report FDA Received Date04/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age55 YR
-
-