Date of event is unknown as it was not provided, best estimate date is (b)(6) 2015 to (b)(6) 2018.(b)(6).The system was evaluated by a field service engineer.All modes of operation and calibrations were verified and no failures were detected.In addition, the field service engineer has followed 15 surgeries and during all the procedures the physician did not complain about the equipment performance.The unit complied with all factory settings.On (b)(6) the field service engineer has followed 15 surgeries and during all the procedures the physician did not complain about the equipment performance.A review of records related to the device including complaint trending, and risk documentation will be performed.Upon completion of this review, if there is any further relevant information obtained, a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc has been submitted.
|
|
The surgery center reported corneal edemas after cataract procedure using a phacoemulsification unit.The surgeon indicated since the purchase of the phacoemulsification unit in year 2015, there has been around 300 cases, in which the surgeon considers around 50 were critical.The surgeon does not have further information about the patient or date of incidents.
|