• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS ITEC PRELOADED 1-PIECE IOL; MONOFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS ITEC PRELOADED 1-PIECE IOL; MONOFOCAL IOLS Back to Search Results
Model Number PCB00
Medical Device Problem Codes Bent (1059); Crack (1135)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 03/22/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If implanted; give date: n/a (not applicable).If explanted; give date: n/a (not applicable).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that intraocular lens (iol) cartridge tip was damaged or bent during handling.There was no patient contact.No additional information provided.
 
Additional Manufacturer Narrative
Device available for evaluation, returned to manufacturer on: 04/24/2018.Device evaluation: the product was received in its original folding carton.The plunger was observed in advanced position and lock.The cartridge was correctly engaged into the device.No assembly error and no defect related to manufacturing process were observed.Visual inspection with microscope magnification was performed.Lack of lubricant material was observed in the device.The lens was stuck at the cartridge tip.The cartridge tip was cracked and the tip of the pushrod came outside through the crack.The reported complaint was verified.However, the condition in which the sample was returned is consistent with a product that was previously handled and prepared for the surgical process.Based on the analysis there is no indication of product quality deficiency.Manufacturing record review: manufacturing record review of the production order and related document revealed that the product was manufactured and released according to specification.There was no discrepancy found during the review.A search revealed that no additional complaint was received from this production order.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the product.Conclusion: based on the investigation results, there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS ITEC PRELOADED 1-PIECE IOL
Common Device Name
MONOFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
santa ana CA 92705
MDR Report Key7449858
Report Number2648035-2018-00585
Device Sequence Number1144406
Product Code HQL
UDI-Device Identifier05050474558144
UDI-Public(01)05050474558144(17)200728
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date07/28/2020
Device Model NumberPCB00
Device Catalogue NumberPCB0000140
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/24/2018
Initial Date Received by Manufacturer 03/23/2018
Supplement Date Received by Manufacturer04/27/2018
Initial Report FDA Received Date04/22/2018
Supplement Report FDA Received Date05/25/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-