• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC KOPANS MODIFIED BREAST LESION LOCALIZATION NEEDLE; MIJ NEEDLE, TUMOR LOCALIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC KOPANS MODIFIED BREAST LESION LOCALIZATION NEEDLE; MIJ NEEDLE, TUMOR LOCALIZATION Back to Search Results
Catalog Number DKBL-20-5.0-A
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/05/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Pma/510(k) #: exempt.(b)(4).This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported during a breast localization, a kopans modified breast lesion localization needle snapped when the surgeon manipulated it after it was placed.As reported, the patient did not experience any adverse effects or require any additional procedures due to this occurrence.
 
Additional Manufacturer Narrative
Investigation - evaluation: a review of the complaint history, drawings, dimensional verification, device history record, instructions for use (ifu), manufacturing instructions, specifications, quality control, and visual inspection of the returned device was conducted during the investigation.We received only one small portion of the kopans breast localization needle back for investigation.Visual examination confirms damage to the surface of the needle.The distal loop appears to be in a damaged condition.Both proximal and distal areas exiting out of the noted cannula appears to be in a damaged condition.The proximal separated end appears to be elongated, twisted, and deformed.No biological matter is present throughout the length of the device.A document-based investigation was performed.There is no evidence to suggest the finished product was not made to specifications.Review of the device history record of the finished product shows no nonconforming events that could contribute to this failure mode.A complaint history search revealed that there were no other reported complaints for this lot number.Based on the information provided, examination of the returned product, and the results of our investigation, a definite root cause could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KOPANS MODIFIED BREAST LESION LOCALIZATION NEEDLE
Common Device Name
MIJ NEEDLE, TUMOR LOCALIZATION
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7450060
Report Number1820334-2018-01019
Device Sequence Number13796149
Product Code MIJ
UDI-Device Identifier00827002025868
UDI-Public(01)00827002025868(17)220706(10)8043531
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberDKBL-20-5.0-A
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 04/06/2018
Supplement Date Received by Manufacturer06/04/2018
Initial Report FDA Received Date04/23/2018
Supplement Report FDA Received Date06/12/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-