| Model Number |
TRUETRACK |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
03/29/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(manufacturer narrative = t, corrected data = f) (b)(4).Returned meter evaluated with no defect found.Test strip vial returned with one strip only - unable to test.Most likely underlying root cause: mlc-20-user's test strip had poor storage.Test strip (b)(4).Note: manufacturer contacted customer on 4/9/2018 in a follow-up call in order to ensure that the replacement products resolved the initial concern; customer stated that the replacement product is working to their satisfaction and they have not had any medical intervention since the last call.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for low blood glucose test results.The customer is concerned with tests results from non-fasting results obtained of 70, 50 and 51 mg/dl.The customer's expected non-fasting blood glucose test result range is 110 - 125 mg/dl; expected fasting blood glucose test result range is 80 - 105 mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call on (b)(6) 2018, a back to back blood test was not performed by the customer.The product was not stored according to specification - customer stated that he left the test strips in the car trunk overnight (8 - 9 hours) and the temperature was below 30 degrees.Customer states that since, the meter has been giving him low results.The test strip lot manufacturer's expiration date is 04/30/2019 and open vial date is (b)(6) 2018.Test strips are not impacted by recall.The meter memory was reviewed for previous test result history: (b)(6).
|
| |
|
Search Alerts/Recalls
|