| Catalog Number |
962323 |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Codes |
Pain (1994); Not Applicable (3189)
|
| Date of Event |
03/27/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
The patient was revised to address instability and pain.Increased thickness of tibial insert.Doi: (b)(6) 2017 dor: (b)(6) 2018 left knee.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary.No device associated with this report was received for examination.If information is obtained that was not available for the initial medwatch , a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|