MAUDE Adverse Event Report: BARD ACCESS SYSTEMS DOT PROV 4F DL FULL TLS; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS DOT PROV 4F DL FULL TLS; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Model Number N/A
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The manufacturer has received the sample and will evaluate.Results are expected soon.A lot history review (lhr) review is not possible, as no manufacturing lot number has been provided by the complainant.
 
Event or Problem Description
The facility reported to the sales rep that a patient's catheter began to leak from a crack at the hub.The catheter was removed and replaced.
 
Additional Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, frequency analysis, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the complaint of a leak is confirmed; however the cause is unknown and an internal investigation,3006260740-12-21-2017-001-c (z#-tbd), is currently ongoing to determine the root cause.One 4 fr dual lumen powerpicc provena was returned for evaluation.An initial visual observation showed use residue throughout the sample, and the ink on the extension legs was observed to be worn and faded.A large split was observed in the extension tubing of the red lumen just within the distal end of the luer connector.Both lumens of the sample were flushed with a 12 ml syringe of water.A leak was observed emanating from the split in the red lumen.No leaks were found in the purple lumen when it was flushed and pressurized and both the purple and red lumens were found to be patent to infusion and aspiration.A microscopic observation revealed the fracture surface of the split in the extension tubing of the red lumen was rough and exhibited cracks/fissures and beach marks, which are indicative of material fatigue.The tubing material of both the red and purple lumens was observed to be attached to the luer connector at irregular points along the tubing¿s circumference, and cracking appeared to have initiated at one or more of these points in both lumens.An internal investigation, 3006260740-12-21-2017-001-c (z#-tbd), is currently ongoing to determine the root cause of this cracking.A lot history review (lhr) review is not possible, as no manufacturing lot number has been provided by the complainant.
 
Event or Problem Description
The facility reported to the sales rep that a patient's catheter began to leak from a crack at the hub.The catheter was removed and replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DOT PROV 4F DL FULL TLS
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key7450678
Report Number3006260740-2018-00791
Device Sequence Number806630
Product Code LJS
UDI-Device Identifier00801741028625
UDI-Public(01)00801741028625
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K162443
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Health Professional
Remedial Action Other,Recall
Type of Report Initial,Followup
Report Date (Section B) 05/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue NumberS9274108
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/20/2018
Event Location Hospital
Initial Date Received by Manufacturer 04/20/2018
Supplement Date Received by Manufacturer04/26/2018
Initial Report FDA Received Date04/23/2018
Supplement Report FDA Received Date05/08/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-