| Model Number |
TRUETRACK |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/02/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(manufacturer narrative = t, corrected data = f) (b)(4).Returned meter evaluated with no defect found.Test strips not returned for evaluation.Most likely underlying root cause : mlc-55 user had an inaccurate reference: competitor's meter: the end user is comparing results obtained from trividia's bgm system to the results from a competitor's bgm system.Test strip (b)(4).Note: manufacturer contacted customer on 4/9/2018 in a follow-up call in order to ensure that the replacement products resolved the initial concern; customer stated that the replacement product is working to their satisfaction and they have not had any medical intervention since the last call.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for high blood glucose test results.The customer is concerned with tests results from results obtained of 223, 369, 174, 416 and 381 mg/dl.The customer's expected fasting blood glucose test result range is 80-150 mg/dl.Customer stated the truetrack meter was reading higher than his other meter; actual meter to meter comparison results were not provided.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.The customer declined to perform a back to back blood test during the call on (b)(6) 2018.The product is stored according to specification in the bedroom.The test strip lot manufacturer's expiration date is 05/31/2020 and open vial date is (b)(6) 2018.The meter memory was reviewed for previous test result history: (b)(6).
|
| |
|
Search Alerts/Recalls
|