• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NEUROSURGERY STRATA II VALVE, SMALL; SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NEUROSURGERY STRATA II VALVE, SMALL; SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS Back to Search Results
Model Number 42856
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 04/19/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that during the implant procedure, the physician found that the valve was leaking.The valve was replaced and there was no injury to the patient.
 
Additional Manufacturer Narrative
The returned valve was patent.The valve met the requirements for siphon, reflux, pressure-flow, pre-implantation testing.There was proteinaceous debris noted within the interior and exterior of the valve.The valve did not meet the requirements for leak testing due to a tear in the top of the delta chamber.It is unknown how or when this damage occurred.The ifu caution that ¿improper use of instruments in the handling or implantation of shunt products may result in the cutting, slitting, or crushing of components.¿ the ifu also caution that ¿care should be taken in handling the valves as silicone has a low cut and tear resistance.All valves are 100% tested at the time of manufacture.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STRATA II VALVE, SMALL
Common Device Name
SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS
Manufacturer (Section D)
MEDTRONIC NEUROSURGERY
125 cremona drive
goleta CA 93117
MDR Report Key7451558
Report Number2021898-2018-00209
Device Sequence Number13484888
Product Code JXG
UDI-Device Identifier00613994804631
UDI-Public00613994804631
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K042465
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/29/2020
Device Model Number42856
Device Catalogue Number42856
Device Lot NumberE30434
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/14/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/19/2018
Supplement Date Received by Manufacturer05/18/2018
Initial Report FDA Received Date04/23/2018
Supplement Report FDA Received Date05/18/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age4 YR
-
-