Catalog Number DYND11552
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 04/09/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
It was reported that the foley catheter was used to provide a method of re-infusing high ostomy output into a lower mucous fistula (to maintain the patency, function and encourage maturation of the distal bowel, refeeding of stoma losses from a high ostomy into a lower mucous fistula may be considered).Reportedly, upon insertion to the patient, air (not saline) approximately 1.5-2 ml was used to inflate the balloon.It was reported that later on, the foley catheter was noted to be out from the patient with the balloon no longer inflated.There was no serious injury reported related to the incident.The reporting facility stated that this is an off-label use of the foley catheter and concluded that this is user-error.Despite good faith efforts to obtain additional information, the reporting facility was unable or unwilling to provide any further patient, product, or procedural details.The sample was not returned for evaluation.Due to the reported incident and in abundance of caution, this medwatch is being filed.If additional relevant information becomes available, a supplemental medwatch will be filed.
 
Event or Problem Description
It was reported that the foley catheter balloon, which was filled with air, would not hold inflation while in the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
CATHETER,FOLEY,100%SILICONE,6FR,1.5ML,LF
Manufacturer (Section D)
MEDLINE INDUSTRIES INC.
three lakes drive
northfield IL 60093
MDR Report Key7451690
Report Number1417592-2018-00030
Device Sequence Number16384442
Product Code EZL
UDI-Device Identifier10884389034415
UDI-Public10884389034415
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Catalogue NumberDYND11552
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/09/2018
Initial Report FDA Received Date04/23/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-