(b)(6).This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.During repair, it was determined that the reported condition was confirmed.The assignable root cause was determined to be due to wear from normal use and servicing.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|
It was reported from the (b)(6) that during service and evaluation, it was determined that the attachment device had temperature above specification, a worn thimble and heated up internally.It was further determined that the device failed pretest for temperature assessment, vibration, thimble lock operation and visual assessment.It was noted in the service order that the device had a defect.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|