• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number MT22495-BLU
Device Problem Inappropriate or Unexpected Reset (2959)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/03/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
Dexcom was made aware on (b)(4) 2018, that on (b)(6) 2018, the receiver initialized without a manual restart.No additional event or patient information is available.The device has been received for investigation.A follow-up will be submitted upon completion of device investigation.
 
Manufacturer Narrative
(b)(4).
 
Event Description
The receiver was returned for evaluation.An external visual inspection was performed and passed.A receiver charge and boot was performed and passed.A receiver download was performed and a screen recoverable error alarm was observed.A functional test was performed and passed.Confirmation of the complaint was confirmed.The probable cause could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
Type of Device
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence dr.
san diego CA 92121
Manufacturer Contact
kipp durbin
6340 sequence dr.
san diego, CA 92121
8582000200
MDR Report Key7451923
MDR Text Key106207211
Report Number3004753838-2018-043101
Device Sequence Number1
Product Code MDS
UDI-Device Identifier30386270000171
UDI-Public30386270000171
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P120005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMT22495-BLU
Device Catalogue NumberSTR-DR-BLU
Device Lot Number5229871
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/20/2018
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/03/2018
Initial Date FDA Received04/23/2018
Supplement Dates Manufacturer Received04/24/2018
Supplement Dates FDA Received05/17/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/06/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age39 YR
-
-