| Model Number |
ZXR00 |
| Medical Device Problem Code |
Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Age/date of birth: unknown, information not provided.Gender/sex: unknown, information not provided.Date of event: unknown, not provided, but the best estimate date is between (b)(6) 2018 and (b)(6) 2018.(b)(4).All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a zxr00 25.5 diopter lens was implanted in the patient's operative eye on (b)(6) 2018.It was later explanted on (b)(6) 2018 due to needing a diopter change.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Device available for evaluation? yes.Returned to manufacturer on: 4/26/2018.Device returned to manufacturer? yes.Device evaluation: the intraocular lens (iol) was returned to the manufacturer.Visual inspection using a microscope was performed and it can be seen that the iol is damaged on the optic and haptic, also it can be seen that the product is being cut in, most probably to make explant possible.The complaint cannot be confirmed.Manufacturing record review: manufacturing records were reviewed and the device was manufactured according to specifications.A complaint history search revealed no similar complaints were received for this production order number to date.Based on the results of the investigation, no product deficiency was identified.All pertinent information available to (b)(4) has been submitted.
|
| |
|