U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN MFG SOLUTIONS S.A. ARGYLE; CATHETER, PERITONEAL, LONG-TERM INDWELLING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN MFG SOLUTIONS S.A. ARGYLE; CATHETER, PERITONEAL, LONG-TERM INDWELLING Back to Search Results
Model Number 8888413815
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/23/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, post laparoscopic installation of catheter in a filtration, when the patient changed position they noticed that the dressing became wet and there was a leak.They still continued with the therapy for several days because of filtration reasons.They tested the catheter and they certified that there was a fissure approximately 1 cm after the second cuff near the radiopaque line.They scheduled for another laparoscopic procedure to remove the catheter, and install a new one to resolve the case.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARGYLE
Common Device Name
CATHETER, PERITONEAL, LONG-TERM INDWELLING
Manufacturer (Section D)
COVIDIEN MFG SOLUTIONS S.A.
edificio b20, calle #2
alajuela 20101
MDR Report Key7452490
Report Number3009211636-2018-00127
Device Sequence Number13716153
Product Code FJS
UDI-Device Identifier10884521004627
UDI-Public10884521004627
Combination Product (Y/N)N
Initial Reporter CountryCI
PMA/510(K) Number
K862046
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Health Professional
Type of Report Initial
Report Date (Section B) 04/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/08/2021
Device Model Number8888413815
Device Catalogue Number8888413815
Device Lot Number1620900134
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/09/2018
Initial Report FDA Received Date04/23/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age47 YR
-
-