| Model Number |
3L81-22 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/03/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The customer stated that an initial creatinine result of 2.3 mg/dl was generated on the architect c4000 analyzer.The result was questioned by the physician.A new sample was obtained and a result of 1.0 mg/dl was generated.The customer repeated both samples and results of 1.0 and 1.1 mg/dl were generated.There was no report of impact to patient management.
|
| |
|
Additional Manufacturer Narrative
|
|
After further evaluation, the customer's issue was is not associated with the clinical chemistry creatinine reagent list number 03l81-22 lot 06816un17.The suspect device is the architect c4000 analyzer and all further information will be documented under manufacturer report number 1628664-2018-00157.
|
| |
|
Search Alerts/Recalls
|