• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - CONTROLLER AC ADAPTER; VENTRICULAR (ASSISST) BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - CONTROLLER AC ADAPTER; VENTRICULAR (ASSISST) BYPASS Back to Search Results
Model Number 1430US
Medical Device Problem Code Power Problem (3010)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/29/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This device is used for treatment not diagnosis.The ventricular assist system is indicated for use as a bridge to cardiac transplantation and destination therapy in patients who are at risk of death from refractory end-stage left ventricular heart failure.The system is designed for in-hospital and out-of-hospital settings.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that post-operative, the controller ac (cac) adapter did not provide power.The cac adapter was connected to the controller, but no green light was seen.Connections were checked and a different outlet was tried, but no power was supplied by the cac adapter.The cac adapter will be exchanged.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product event summary: the controller ac adapter was returned for evaluation.Various analyses were conducted and reviewed in order to evaluate the performance of the device in relation to the reported event.Failure analysis of the returned device revealed that the controller ac adapter passed visual examination and functional testing.The controller ac adapter was able to adequately provide power to a test controller.As a result, the reported event could not be confirmed.A possible root cause of the reported event can be attributed to a marginal connection between the controller ac adapter and the electrical outlet.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The cac adapter was exchanged.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTWARE VENTRICULAR ASSIST SYSTEM - CONTROLLER AC ADAPTER
Common Device Name
VENTRICULAR (ASSISST) BYPASS
Manufacturer (Section D)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
MDR Report Key7453003
Report Number3007042319-2018-01562
Device Sequence Number13727410
Product Code DSQ
UDI-Device Identifier00888707002738
UDI-Public00888707002738
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
P100047
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 09/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date01/16/2019
Device Model Number1430US
Device Catalogue Number1430US
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/06/2018
Initial Date Received by Manufacturer 03/29/2018
Supplement Date Received by Manufacturer09/13/2018
Initial Report FDA Received Date04/23/2018
Supplement Report FDA Received Date09/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age66 YR
Patient Weight56
-
-