| Model Number |
ZCT150 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
03/27/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a zct150 22.5 diopter lens was implanted into the patient's left eye (os) on (b)(6) 2018.It was later explanted on (b)(6) 2018 due to an unexpected post-operative refraction (distance vision 20/50).There was no incision enlargement, no vitrectomy, and no sutures used.Reportedly, there was no patient injury and the patient is doing fine post-operatively.The replacement lens was zcb00 21.5 diopter lens.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the product was not returned to the manufacturing site.Therefore, product testing could not be performed and the customer¿s reported complaint could not be verified.Manufacturing records review: the manufacturing records for the product was reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no other complaints have been received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Search Alerts/Recalls
|