KISCO INTERNATIONAL L-VARLOCK; LUMBAR POSTERIOR CAGE Back to Search Results
Model Number 10 MM X 9 MM
Medical Device Problem Code Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/21/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device history record of the batch (160100) was reviewed and no deviation was detected.The medical device complied with the specification.The cage was expanded to the maximum (maybe up to or more than 10 turns) and the stresses applied to the cage probably caused the screw breakage.The surgical technique states that two cages must be implanted in the intervertebral space.In this case, just one cage, implanted using an oblique approach was implanted.We are waiting for the x-rays to confirm the screw breakage of the cage and to determine the root cause.Submission of a medical device report and the fda's release of that information is not an admission that a product, user facility, importer, distributor, manufacturer, or medical personnel caused or contributed to the event.
 
Event or Problem Description
During surgery the surgeon expanded (maybe up to more than 10 turns) l-varlock lumbar cage.The surgeon reported that he felt the expansion screw broke since the screwdriver kept spinning.To date we do not have visual confirmation that the screw broke after several requests to the surgeon.Submission of a medical device report and the fda's release of that information is not an admission that a product, user facility, importer, distributor, manufacturer, or medical personnel caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
L-VARLOCK
Common Device Name
LUMBAR POSTERIOR CAGE
Manufacturer (Section D)
KISCO INTERNATIONAL
parc technologique
2, place berthe morisot
saint priest, 69800
FR  69800
MDR Report Key7453331
Report Number3005440746-2018-00003
Device Sequence Number19072420
Product Code MAX
UDI-Device Identifier03700406328018
UDI-Public(01)03700406328018
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K080537
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source distributor
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Model Number10 MM X 9 MM
Device Catalogue NumberVAL-CA 10-09
Device Lot Number160100
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/21/2018
Initial Report FDA Received Date04/23/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age60 YR
-
-