| Model Number |
MMT-523LNAB |
| Health Effect - Clinical Code |
Hyperglycemia (1905)
|
| Date of Event |
03/17/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that a piece of reservoir lip ring was broken off.The customer¿s blood glucose level was little bit over 200 mg/dl.The customer was assisted with the troubleshooting.The customer stated that the reservoir compartment was coming loose.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
A test reservoir was installed and did not lock in place due to completely broken reservoir tube lip.Unable to perform rewind test, basic occlusion test, occlusion test, prime test, excessive no delivery test and displacement test due to completely broken reservoir tube lip.Device had missing reservoir tube o-ring.
|
| |
|