• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC, INC. INTERSTIM ICON; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC. INTERSTIM ICON; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE Back to Search Results
Model Number 3037
Medical Device Problem Code Device Displays Incorrect Message (2591)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/26/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from a healthcare provider (hcp) regarding a patient with an implantable neurostimulator (ins) for urinary dysfunction/sacral nerve stimulation and gastrointestinal/pelvic floor.The hcp was encountering the poor communication screen and indicated that it took 10 minutes to decrease from 1.8 to 1.4.The issue appeared to be that the patient was not holding the patient programmer (pp) on the skin pressing the minus button.The caller indicated that the patient was in the hospital about to have surgery, but the type of surgery or if the surgery was related to the patient's therapy was not reported.Patient status is unknown at the time of this report.There were no further complication reported or anticipated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERSTIM ICON
Common Device Name
STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
MDR Report Key7453690
Report Number2182208-2018-00640
Device Sequence Number19034249
Product Code EZW
UDI-Device Identifier00643169512955
UDI-Public00643169512955
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P970004
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 04/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number3037
Device Catalogue Number3037
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/26/2018
Initial Report FDA Received Date04/24/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age60 YR
-
-