• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE Back to Search Results
Model Number 8637-40
Device Problems Volume Accuracy Problem (1675); Insufficient Flow or Under Infusion (2182)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/11/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a healthcare provider via a clinical study regarding a patient receiving hydromorphone (2.0 mg/ml at 0.8999 mg/day), compounded baclofen (300.0 mcg/ml at 134.98 mcg/day), and bupivacaine (30.0 mg/ml at 13.498 mg/day) via an implanted pump.The indication for use was spinal pain.It was reported during a pump refill, on (b)(6) 2018, a reservoir volume discrepancy was noted in which the expected reservoir volume was 4.8 ml and the actual reservoir volume was 21 ml.The device diagnosis was pump reservoir volume discrepancy.It was indicated the event was related to the device or therapy.No associated signs or symptoms were noted.No further diagnostics or interventions were performed.The issue was unresolved with no further action planned.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information received from healthcare professional (hcp) via a clinical study indicated that the cause of the volume discrepancy was not determined.The patient's baseline weight was no measured.No further complications were reported/anticipated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCHROMED II
Type of Device
PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key7454010
MDR Text Key106398017
Report Number3004209178-2018-08653
Device Sequence Number1
Product Code LKK
UDI-Device Identifier00643169508156
UDI-Public00643169508156
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,study
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/24/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/14/2018
Device Model Number8637-40
Device Catalogue Number8637-40
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/25/2018
Date Device Manufactured02/15/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age60 YR
-
-