• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVIS & GECK CARIBE LTD NOVAFIL; SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVIS & GECK CARIBE LTD NOVAFIL; SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE Back to Search Results
Model Number 8886445251
Medical Device Problem Codes Detachment Of Device Component (1104); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 04/11/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, the needle detached from the suture.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of four device.The visual inspection of the open sample noted two partial sutures and one needle.The needle was observed to be detached from the sutures.Upon microscopic inspection of the needle, normal needle crimp and swage marks were observed on the needle.The retainer was inspected with no abnormalities.The visual inspection and functional evaluation of the sealed samples had acceptable results.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all manufacturers¿ quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.Should new information become available, the file will be re-opened, and the investigation summary amended as appropriate.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during gastric sleeve procedure, the needle broke away from the suture at the needle point attachment/ clamped join between the two.The separated parts were expelled from the patient and a new suture packet was opened to complete the procedure.There was no patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NOVAFIL
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DO  0101
MDR Report Key7454146
Report Number9612501-2018-00826
Device Sequence Number1264388
Product Code GAT
UDI-Device Identifier10884521100091
UDI-Public10884521100091
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K990952
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 06/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2021
Device Model Number8886445251
Device Catalogue Number8886445251
Device Lot NumberD6H0701X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/26/2018
Initial Date Received by Manufacturer 04/11/2018
Supplement Date Received by Manufacturer05/30/2018
Initial Report FDA Received Date04/24/2018
Supplement Report FDA Received Date06/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age38 YR
-
-