Date of event is unknown.Additional device product code: hwc.Date of implant is an unknown date in 2016.Device is not expected to be returned for manufacturer review/investigation.(b)(4) used to capture additional medical/surgical intervention required.Device history records review was completed for part# 456.306, lot# 7794143.Manufacturing location: (b)(4), manufacturing date: sep 18, 2014.A review of the device history record revealed no complaint related anomalies.The device history record shows this lot of 11.0mm ti helical blade 105mm nonsterile product was processed through the normal manufacturing and inspection operations with no rework nor nonconformities noted.This lot met all dimensional, visual and packaging criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.A review of the raw material device history record revealed this lot met all specifications with no non-conformance noted.This raw material lot met all dimensional and visual criteria at the time of manufacture with no issues documented during the manufacture that would contribute to this complaint condition.The investigation could not be completed; no conclusion could be drawn, as no product was received.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
It was reported that the patient was implanted with a trochanteric femoral nail (tfn) in 2016.The patient suffered from various illness and is advanced in age.Due to infection, the patient underwent a removal surgery on (b)(6) 2018.It was reported that the tfn nail, helical blade and a 5.0 locking screw were removed and nothing was replaced.It was also reported that at the end of the procedure there was a complete irrigation of the infected area which was the entire left leg of the patient.There were laboratory culture tests taken due to infection.There was no surgical delay.The patient outcome is unknown.Surgery was completed successfully with no delay.This report is for one (1) 11.0mm ti helical blade 105mm.This is report 2 of 3 for (b)(4).
|