Model Number C20
Device Problem Leak/Splash (1354)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that the anesthesia workstation failed the leakage test during system check out.There was no patient connected to the device during the event.(b)(4).
 
Manufacturer Narrative
The company service engineer investigated the anesthesia workstation at the hospital and concluded that the manual/automatic ventilation valve membrane in the patient cassette and the actuator manual/automatic ventilation valve had been damaged.The damaged parts were replaced but were not returned for investigation.Images of the damaged parts were received and our conclusion is that the damages were caused by attempting to remove the patient cassette while the manual/automatic ventilation valve actuator was in its upper position, i.E excessive or inadequate physical force had been applied to the patient cassette.The system check out in the supplied log failed due to the leakage caused by the damaged parts.Getinge usa sales, llc (importer) is submitting this report on behalf of maquet critical care ab (manufacturer).Ref.Exemption #: e2018003.(b)(4).
 
Event Description
(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLOW-I C20
Type of Device
GAS-MACHINE, ANESTHESIA
Manufacturer (Section D)
MAQUET CRITICAL CARE AB
solna
SW 
Manufacturer (Section G)
MAGNUS LINDQVIST
maquet critical care ab
röntgenvägen 2, se-171 54
solna
SW  
Manufacturer Contact
maquet critical care ab
röntgenvägen 2, se-171 54
solna 
MDR Report Key7454491
MDR Text Key106722166
Report Number8010042-2018-00200
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
PMA/PMN Number
K160665
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberC20
Device Catalogue Number6677200
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/02/2018
Initial Date FDA Received04/24/2018
Supplement Dates Manufacturer Received04/02/2018
Supplement Dates FDA Received05/17/2018
Date Device Manufactured06/18/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-