| Model Number |
LS1520 |
| Medical Device Problem Codes |
Device Stops Intermittently (1599); Difficult to Open or Close (2921)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
04/05/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a procedure, the device¿s jaws closed and could not activate or open while being applied to tissue.The jaws eventually opened and they grabbed a new instrument.There was no patient injury or blood loss.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: one device was received for evaluation.This device had been used in the treatment or diagnosis of a patient.A review of the lot number reported indicates that the product was within the assigned expiration date at the time of the reported incident.The returned product met specification as received.Visual inspection found no defects.The reported condition was not confirmed.The investigation found the jaws opened and closed properly when the handle was retracted and released.The device was activated ten times on a saline soaked towel with satisfactory results.All seal cycles were completed satisfactorily and end tones were heard indicating completed activation cycles.The investigation found the device to function normally and within specifications.If information is provided in the future, a supplemental report will be issued.
|
| |
|