• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN MEDICAL PRODUCTS LIGASURE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN MEDICAL PRODUCTS LIGASURE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number LS1520
Medical Device Problem Codes Device Stops Intermittently (1599); Difficult to Open or Close (2921)
Health Effect - Clinical Code No Information (3190)
Date of Event 04/05/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a procedure, the device¿s jaws closed and could not activate or open while being applied to tissue.The jaws eventually opened and they grabbed a new instrument.There was no patient injury or blood loss.
 
Additional Manufacturer Narrative
Evaluation summary: one device was received for evaluation.This device had been used in the treatment or diagnosis of a patient.A review of the lot number reported indicates that the product was within the assigned expiration date at the time of the reported incident.The returned product met specification as received.Visual inspection found no defects.The reported condition was not confirmed.The investigation found the jaws opened and closed properly when the handle was retracted and released.The device was activated ten times on a saline soaked towel with satisfactory results.All seal cycles were completed satisfactorily and end tones were heard indicating completed activation cycles.The investigation found the device to function normally and within specifications.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGASURE
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
COVIDIEN MEDICAL PRODUCTS
building 10- no 789 puxing roa
shanghai 20111 4
CN  201114
MDR Report Key7454647
Report Number3006451981-2018-00297
Device Sequence Number1373564
Product Code GEI
UDI-Device Identifier10884524002187
UDI-Public10884524002187
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K043273
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 05/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/12/2021
Device Model NumberLS1520
Device Catalogue NumberLS1520
Device Lot NumberS6KB003PX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/14/2018
Initial Date Received by Manufacturer 04/11/2018
Supplement Date Received by Manufacturer05/15/2018
Initial Report FDA Received Date04/24/2018
Supplement Report FDA Received Date05/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-