| Catalog Number |
SC60 |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/22/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Batch # (b)(4).Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date, a supplemental medwatch will be sent.The device history records were reviewed and the manufacturing criteria were met prior to the release of this batch/lot.
|
| |
|
Event or Problem Description
|
|
It was reported that the staples malformed after firing.During the vats lobectomy, after fired with ecr60b, when severing the lobe of lung, found the staples malformed.Another device was used to complete the surgery.There were no adverse consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Batch # p55m2e.Device evaluation: the analysis results found that one sc60 device was returned with no apparent damage and with no cartridge reload present.The device was tested for functionality with a test cartridge reload and achieved its complete 3-stroke firing sequence without any difficulties.The staple line and cut line were complete and the staples were noted to have the proper b-form shape.The event described could not be confirmed as the device performed without any difficulties noted.The batch history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the manufacturing process.
|
| |
|