MAUDE Adverse Event Report: COOK INC ADVANCE 35 LP LOW PROFILE BALLOON CATHETER; DQY CATHETER, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC ADVANCE 35 LP LOW PROFILE BALLOON CATHETER; DQY CATHETER, PERCUTANEOUS Back to Search Results
Model Number G52270
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/06/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported that during angioplasty of the left superficial femoral artery, the advance 35 lp low profile balloon catheter ruptured at 10 atmospheres (atm) inflation pressure.There was no vessel angulation or calcification.Nonionic contrast with a ratio of 7ml contrast to 3ml 0.9% saline solution was used.Another device was used to successfully complete the procedure.Per the complainant, the patient remained stable through the end of the procedure.According to the initial reporter, a section of the device did not remain inside the patient's body.The patient did not experience any adverse effects due to this occurrence.
 
Additional Manufacturer Narrative
Investigation ¿ evaluation: a review of the complaint history, device history record, documentation, drawing, instructions for use, quality control, and specifications of the device was conducted during the investigation.The complaint device was not returned; therefore, no physical examinations could be performed.However, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record shows no nonconforming events which could contribute to this failure mode.It should be noted there were no other reported complaints for this lot number.Based on the information provided, no product returned and the results of our investigation, a definitive root cause could not be determined.Measures are being conducted to address this failure mode.We will continue to monitor for similar complaints.Per the quality engineering risk assessment no further action is required.
 
Additional Manufacturer Narrative
A photograph of the complaint device was returned by the customer, and a photographic evaluation was performed as a part of the complaint investigation.From the photograph, it is difficult to determine if there is a balloon rupture or not.No further information was provided.
 
Event or Problem Description
Additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCE 35 LP LOW PROFILE BALLOON CATHETER
Common Device Name
DQY CATHETER, PERCUTANEOUS
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7454715
Report Number1820334-2018-01045
Device Sequence Number11171870
Product Code DQY
UDI-Device Identifier10827002522708
UDI-Public(01)10827002522708(17)200223(10)7701000
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG52270
Device Catalogue NumberPTA5-35-135-6-14.0
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/09/2018
Supplement Date Received by Manufacturer05/03/2018
07/16/2018
Initial Report FDA Received Date04/24/2018
Supplement Report FDA Received Date05/18/2018
07/25/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age77 YR
-
-