| Model Number |
CRPLUS |
| Medical Device Problem Codes |
Failure to Charge (1085); Failure to Deliver Energy (1211)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
03/09/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Physio-control evaluated the customer's device and verified the reported issue.Physio observed that the device can not recognize a shockable patient rhythm to deliver defibrillation energy and does not recognize a patient connection had been made.Physio removed the analog pcb assembly for further evaluation.The cause of the reported issue was determined to be due to capacitor, designator c157.C157, leg 3 to leg 2, is electrically shorted from process residue and caused the device to not be able to charge and shock.The customer received a replacement device.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report a non-critical issue with their device related to the service wrench appearing on the readiness display.There was no patient use associated with the reported event.Upon evaluation of the customer's device, physio-control observed that their device would not charge and shock.As a result, defibrillation energy would not be delivered, if it were necessary.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report a non-critical issue with their device related to the service wrench appearing on the readiness display.There was no patient use associated with the reported event.Upon evaluation of the customer's device, physio-control observed that their device would not charge and shock.As a result, defibrillation energy would not be delivered, if it were necessary.
|
| |
|
Additional Manufacturer Narrative
|
|
Gtin/device identifier (di) number ¿ (b)(4).
|
| |
|