• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MPRI ATTAIN ABILITY; DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MPRI ATTAIN ABILITY; DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE Back to Search Results
Model Number 419678
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code Undesired Nerve Stimulation (1980)
Date of Event 04/05/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
C6tr01 crtp implanted: (b)(6) 2018.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient experienced phrenic nerve stimulation.The patient's left ventricular (lv) lead was explanted and replaced.No further patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product event summary: the full lead was returned, analyzed, and no anomalies were found.Visual analysis of the lead indicated apparent explant damage.Visual analysis of the lead indicated the lead was stretched.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTAIN ABILITY
Common Device Name
DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
MDR Report Key7456011
Report Number2649622-2018-05656
Device Sequence Number1083528
Product Code OJX
UDI-Device Identifier00643169860025
UDI-Public00643169860025
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P080006
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date06/02/2019
Device Model Number419678
Device Catalogue Number419678
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/11/2018
Initial Date Received by Manufacturer 04/05/2018
Supplement Date Received by Manufacturer06/04/2018
Initial Report FDA Received Date04/24/2018
Supplement Report FDA Received Date06/12/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
4068-52
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age77 YR
-
-