Model Number 16-02-80 |
Device Problem
Device Displays Incorrect Message (2591)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 03/29/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
The heater-cooler 16-02-80 is not distributed in the usa, but it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k052601).Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.Device not returned.
|
|
Event Description
|
Livanova (b)(4) received a report that a heater-cooler system 3t displayed an error message during in-service.There was no patient involvement.
|
|
Manufacturer Narrative
|
A livanova field service representative was dispatched to the facility to investigate.The service representative was able to confirm the reported issue and traced the failure to a faulty pump.The pump was replaced to resolve the reported issue.Subsequent functional verification testing was completed without further issues and the unit was returned to service.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
See initial.
|
|
Manufacturer Narrative
|
The defective pump was returned for further evaluation.During the evaluation the reported issue could be reproduced.During the evaluation the bearings of the pump were found to be rusty; this led to the reported malfunction.
|
|