• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHYSIO-CONTROL, INC. - 3015876 LIFEPAK CR® PLUS DEFIBRILLATOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHYSIO-CONTROL, INC. - 3015876 LIFEPAK CR® PLUS DEFIBRILLATOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Model Number CRPLUS
Medical Device Problem Code Power Problem (3010)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 04/02/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Physio-control evaluated the customer's device and verified the reported issue.Physio also observed that the device would not remain powered on to charge and shock.Physio determined that the cause of the reported issue was that the device had an excessive amount of on-time which depleted the internal hybrid layer capacitor (hlc) batteries and prevented the device from remaining powered on to charge and shock.The customer received a replacement device.
 
Event or Problem Description
The customer contacted physio-control to report that their device was showing all three icons (attention, charge-pak and service wrench).With all three icons illuminated, the device may not have sufficient power available to deliver effective defibrillation therapy to a patient, if it was needed.There was no patient use associated with the reported event.
 
Additional Manufacturer Narrative
Gtin/device identifier (di) number ¿(b)(4) section d4 of the initial medwatch report indicates: model # ¿ blank section d4 of the initial medwatch report should indicate: model # ¿ crplus.
 
Event or Problem Description
The customer contacted physio-control to report that their device was showing all three icons (attention, charge-pak and service wrench).With all three icons illuminated, the device may not have sufficient power available to deliver effective defibrillation therapy to a patient, if it was needed.There was no patient use associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK CR® PLUS DEFIBRILLATOR
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
MDR Report Key7456671
Report Number0003015876-2018-00662
Device Sequence Number13693790
Product Code MKJ
UDI-Device Identifier00885074440156
UDI-Public00885074440156
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
PMA/510(K) Number
K033275
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberCRPLUS
Device Catalogue Number99403-000157
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/16/2018
Initial Date Received by Manufacturer 04/02/2018
Supplement Date Received by Manufacturer02/20/2020
Initial Report FDA Received Date04/24/2018
Supplement Report FDA Received Date02/21/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-